Product Manual

Calibration
Calibration Routine16
16-5
CAL 56 CH2 Level 35 MHz Adjust for same reading
CAL 57 CH2 Level 37·5 MHz Adjust for same reading
CAL 58 CH2 Level 40 MHz Adjust for same reading
CAL 59 CH3 DC offset zero. Adjust for 0 V ±5 mV
CAL 60 CH3 DC offset at + full scale. Adjust for +10 V ±10 mV
CAL 61 CH3 DC offset at full scale. Check for –10 V ±3 %
CAL 62 CH3 Multiplier zero. Adjust for minimum volts ac
CAL 63 CH3 Multiplier offset. Adjust for 0 V ±5 mV
CAL 64 CH3 Waveform offset. Adjust for 0 V ±5 mV
CAL 65 CH3 Output level at full scale Adjust for 10 V ±10 mV
CAL 66 CH3 20 dB attenuator Adjust for 1 V ±1 mV
CAL 67 CH3 40 dB attenuator Adjust for 0·1 V ±·1 mV
CAL 68 CH3 10 dB attenuator Adjust for 2·236 V ac ±10 mV
CAL 69 CH3 Sum offset Adjust for 0 V ±5 mV
CAL 70 CH3 SCM level at full-scale Adjust for 5 V ±5 mV
CAL 71 CH3 AM level at full-scale Adjust for 10 V ±10 mV
CAL 72 CH3 Level 0.1 MHz Note reading
CAL 73 CH3 Level 33 MHz Check reading
CAL 74 CH3 Level 1 MHz Adjust for same reading
CAL 75 CH3 Level 2 MHz Adjust for same reading
CAL 76 CH3 Level 4 MHz Adjust for same reading
CAL 77 CH3 Level 5 MHz Adjust for same reading
CAL 78 CH3 Level 10 MHz Adjust for same reading
CAL 79 CH3 Level 15 MHz Adjust for same reading
CAL 80 CH3 Level 20 MHz Adjust for same reading
CAL 81 CH3 Level 25 MHz Adjust for same reading
CAL 82 CH3 Level 30 MHz Adjust for same reading
CAL 83 CH3 Level 32·5 MHz Adjust for same reading
CAL 84 CH3 Level 35 MHz Adjust for same reading
CAL 85 CH3 Level 37·5 MHz Adjust for same reading
CAL 86 CH3 Level 40 MHz Adjust for same reading
CAL 87 CH4 DC offset zero. Adjust for 0 V ±5 mV
CAL 88 CH4 DC offset at + full scale. Adjust for +10 V ±10 mV
CAL 89 CH4 DC offset at full scale. Check for –10 V ±3 %
CAL 90 CH4 Multiplier zero. Adjust for minimum volts ac
CAL 91 CH4 Multiplier offset. Adjust for 0 V ±5 mV
CAL 92 CH4 Waveform offset. Adjust for 0 V ±5 mV
CAL 93 CH4 Output level at full scale Adjust for 10 V ±10 mV
CAL 94 CH4 20 dB attenuator Adjust for 1 V ±1 mV
CAL 95 CH4 40 dB attenuator Adjust for 0·1 V ±·1 mV
CAL 96 CH4 10 dB attenuator Adjust for 2·236 V ac ±10 mV
CAL 97 CH4 Sum offset Adjust for 0 V ±5 mV
CAL 98 CH4 SCM level at full-scale Adjust for 5 V ±5 mV
CAL 99 CH4 AM level at full-scale Adjust for 10 V ±10 mV
CAL 100 CH4 Level 0.1 MHz Note reading
CAL 101 CH4 Level 33 MHz Check reading
CAL 102 CH4 Level 1 MHz Adjust for same reading
CAL 103 CH4 Level 2 MHz Adjust for same reading
CAL 104 CH4 Level 4 MHz Adjust for same reading
CAL 105 CH4 Level 5 MHz Adjust for same reading
CAL 106 CH4 Level 10 MHz Adjust for same reading
CAL 107 CH4 Level 15 MHz Adjust for same reading
CAL 108 CH4 Level 20 MHz Adjust for same reading
CAL 109 CH4 Level 25 MHz Adjust for same reading
CAL 110 CH4 Level 30 MHz Adjust for same reading