Product Manual

Connections
Rear Panel Connections 3
3-5
Pin 2, 3 and 5 may be used as a conventional RS232 interface with XON/XOFF
handshaking. Pins 7, 8 and 9 are used when the instrument is operated in addressable
RS232 mode. Signal grounds are connected to the instrument ground. The RS232 address
is set from the
remote menu on the UTILITY screen, as described in chapter 14,
System Operations from the Utility Menu.
GPIB (IEEE-488)
The GPIB interface is not isolated; the GPIB signal grounds are connected to the
instrument ground.
The implemented subsets are:
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0, E2.
The GPIB address is set from the
remote menu on the UTILITY screen, as described
chapter 14, System Operations from the Utility Menu.