Product Manual

291, 292, 294
Users Manual
17-22
LOCAL Returns the instrument to local operation and
unlocks the keyboard.
Will not function if LLO is in force.
UCBID? Returns the instrument's address.
Refer to chapter 16, Calibration, for remote calibration commands.
Remote Command Summary
Table 17-1. Remote Command Summary
*
CLS Clear status.
*
ESE <nrf> Set the Standard Event Status Enable Register
to the value of <nrf>.
*
ESE? Returns the value in the Standard Event Status
Enable Register in <nr1> numeric format.
*
ESR? Returns the value in the Standard Event Status
Register in <nr1> numeric format.
*
IDN? Returns the instrument identification.
*
IST? Returns ist local message as defined by
IEEE Std. 488.2.
*
LRN? Returns the complete set up of the instrument
as a hexadecimal character data block
approximately 842 bytes long.
*
OPC Sets the Operation Complete bit (bit 0) in the
Standard Event Status Register.
*
OPC? Query operation complete status.
*
PRE <nrf> Set the Parallel Poll Enable Register to the
value <nrf>.
*
PRE? Returns the value in the Parallel Poll Enable
Register in <nr1> numeric format.
*
RCL <cpd> Recalls the instrument set up contained in store
<cpd>.
*
RST Resets the instrument parameters to their
default values.
*
SAV <cpd> Saves the complete instrument set up to the set-
up file named <cpd>.
*
SRE <nrf> Set the Service Request Enable Register to
<nrf>.
*
SRE? Returns the value of the Service Request
Enable Register in <nr1> numeric format.
*
STB? Returns the value of the Status Byte Register in
<nr1> numeric format.
*
TRG This command is the same as pressing the
MAN TRIG
key.