Electrical Specifications
Specifications are valid for 5-½ digit mode and after at least a half-hour warm-up.
DC Voltage Specifications
Maximum Input ..........................................................1000 V on any range
C
ommon Mode Rejection..........................................120 dB at 50 or 60 Hz r0.1㧑 (1 k: unbalance)
N
ormal Mode Rejection.............................................80 dB at Slow Rate
A
/D Nonlinearity ........................................................15 ppm of range
I
nput Bias Current .....................................................<30 pA at 25 rC
S
ettling Considerations.............................................Measurement settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes
I
nput Characteristics
R
esolution
Range
F
ull-Scale
(5-1/2 Digits)
S
low Medium Fast
Input Impedance
200 mV 199.999 mV 1 PV 10 PV 10 PV
>10 G:
[1]
2 V 1.99999 V 10 PV 100 PV 100 PV
>10 G:
[1]
20 V 19.9999 V 100 PV 1000 PV 1000 PV
10 M:r1 %
200 V 199.999 V 1 mV 10 mV 10 mV
10 M:r1 %
1000 V 1000.00 V 10 mV 100 mV 100 mV
10 M:r1 %
Notes:
[1] At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M:.
U
ncertainty
[1]
90 days 1 year
Range
2
3 °C r 5 °C 23 °C r 5°C
T
emperature Coefficient/°C
Outside 18 – 28 °C
200 mV 0.01 + 0.003 0.015 + 0.004 0.0015 + 0.0005
2 V 0.01 + 0.002 0.015 + 0.003
0.001 + 0.0005
20 V 0.01 + 0.003 0.015 + 0.004
0.0020 + 0.0005
200 V 0.01 + 0.002 0.015 + 0.003
0.0015 + 0.0005
1000 V 0.01 + 0.002 0.015 + 0.003
0.0015 + 0.0005
Notes:
[1] Uncertainty given as r (% of reading + % of range)