Item Brochure
6 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
6 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
U
ncertainty
[1]
90 days 1 year
Range
2
3 °C r 5 °C 23 °C r 5 °C
T
emperature Coefficient/°C
O
utside 18 – 28 °C
200 PA 0.02 + 0.005 0.03 + 0.005 0.003 + 0.001
2 mA 0.015 + 0.005 0.02 + 0.005 0.002 + 0.001
20 mA 0.03 + 0.02 0.04 + 0.02 0.005 + 0.001
200 mA 0.02 + 0.005 0.03 + 0.008 0.005 + 0.001
2 A 0.05 + 0.02 0.08 + 0.02 0.008 + 0.001
10 A 0.18 + 0.01 0.2 + 0.01 0.008 + 0.001
Notes:
[1] Uncertainty given as r (% of reading + % of range)
AC Current
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs
from 1 % to 5 % of range, add an additional error of 0.1 % of range.
I
nput Protection .........................................................Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses
M
easurement Method ...............................................AC-coupled True RMS
S
hunt Resistance .......................................................0.01 : for 2 A and 10 A ranges
1 : for 20 mA and 200 mA
A
C Filter Bandwidth..................................................20 Hz – 100 kHz
M
aximum Crest Factor ..............................................3:1 at Full Scale
A
dditional Crest Factor Errors (<100 Hz) ..............Crest Factor 1-2, 0.05 % of full scale
Crest Factor 2-3, 0.2 % of full scale
Only applies to non-sinusoid signals
I
nput Characteristics
R
esolution
Range
F
ull-Scale
(5-1/2 Digits)
S
low Medium Fast
B
urden Voltage
20 mA 19.9999 mA 0.1 PA 1 PA 1 PA
<0.05 V
200 mA 199.999 mA 1 PA 10 PA 10 PA
<0.5 V
2 A 1.99999 A 10 PA 100 PA 100 PA
<0.1 V
10 A 10.0000 A 100 PA 1 mA 1 mA
<0.5 V
U
ncertainty
[1]
90 days 1 year
Range Frequency
2
3 °C r 5 °C 23 °C r 5 °C
T
emperature
Coefficient/°C
Outside 18 – 28 °C
20 mA 20 Hz - 45Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.005
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
200 mA 20 Hz - 45Hz 0.8 + 0.05 1 + 0. 06 0.015 + 0.005
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
2 A 20 Hz - 45Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.005
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
10 A 20 Hz - 45Hz 1 + 0.1 1.25 + 0.12 0.015 + 0.005
45 Hz - 2 kHz 0.35 + 0.1 0.5 + 0.12 0.015 + 0.005
Notes: