Manual Supplement Manual Title: Part Number: Print Date: Revision/Date: 345 Calibration 3095315 October 2007 Supplement Issue: Issue Date: Page Count: 5 8/13 5 This supplement contains information necessary to ensure the accuracy of the above manual. © 2008-2013 Fluke Corporation. All rights reserved.
345 Calibration Manual Supplement Change #1 On page 26, under Power Check-Stage 1, replace the entire table with: Calibrator Specification (k)W Volts (V) Amps (A) Phase Shift (°) Min Nom Max 19 100 0.00 1820 1900 1980 21 100 0.00 2042 2100 2158 38 100 0.00 3700 3800 3900 44 100 0.00 4.24 4.40 4.56 380 100 0.00 37.00 38.00 39.00 440 100 0.00 42.4 44.0 45.6 Calibrator Specification (k)VA Volts (V) Amps (A) Phase Shift (°) Min Nom Max 19 100 0.
Manual Supplement 345 Calibration On page 27, under Power Check-Stage 2, replace the entire table with: For 345 with Firmware V1.1 Calibrator Specification (k)W Volts (V) Amps (A) Phase Shift (°) Min Nom Max 19.8681 100 73.00 501 581 661 21.9595 100 -73.00 562 642 722 39.2133 100 73.00 1066 1146 1226 46.0105 100 -73.00 1.27 k 1.35 k 1.43 k 397.363 100 73.00 10.83 k 11.62 k 11.96 k 460.105 100 -73.00 12.6 k 13.5 k 13.
345 Calibration Manual Supplement For 345 with Firmware V1.2 or greater Calibrator Specification (k)W Volts (V) Amps (A) Phase Shift (°) Min Nom Max 19.8681 100 73.00 501 581 661 21.9595 100 -73.00 562 642 722 39.2133 100 73.00 1066 1146 1226 46.0105 100 -73.00 1.27 k 1.35 k 1.43 k 397.363 100 73.00 10.83 k 11.62 k 11.96 k 460.105 100 -73.00 12.6 k 13.5 k 13.8 k Calibrator Specification (k)VAR Volts (V) Amps (A) Phase Shift (°) Min Nom Max 19.
345 Calibration Manual Supplement Change #2, 63545 On page 4, add the following to the Symbols table: Conforms to relevant South Korean EMC Standards. Conforms to CAN/CSA-C22.2 No. 61010-1, second edition, including Amendment 1. Change #3, 65704 On page 9, under General Data: Change: Battery Eliminator BE345 Input……110V / 230V, 50/60 Hz Output….15 V dc, 300 mA To: Battery Eliminator GFP151DA-120125B-1 Input……100 V to 240 V, 50/60 Hz, 0.36 A Output….12 V dc, 1.
345 Calibration Manual Supplement On page 9, replace Electrical Safety and EMC sections, with: General Specifications Maximum working voltage in CAT IV areas: Current measurement: ............................... 600 V ac rms or dc between conductor & ground Voltage measurement ................................ 600 V ac rms or dc between either input terminal and ground, or 825 V between energized phase voltages (delta power config.) Maximum working voltage in CAT III areas ....