User's Manual

Performance Verification
Accuracy Performance Tests
4
4-11
Volts AC Accuracy Test
The following procedure tests the accuracy of the volts ac function for both the PAI and
FAI modules. Measurement accuracy applies to all channels, not just the channel used for
the test.
1. Configure Channel 1 for V ac Using the Fluke DAQ software, configure channel 1
for V ac, 300 mV range.
2. Open Spy Open the Spy window under the Configuration dialog.
3. Verify Accuracy Configure the 5700A for the output values below and verify the
measurement is between the minimum and maximum values. Change the channel 1
range as required (see Step 1).
Volts AC Range 5700A Output Minimum Reading Maximum Reading
300 mV 20 mV@1 kHz 0.0197 V 0.0203 V
300 mV 20 mV@100 kHZ 0.0185 V 0.0215 V
300 mV 300 mV@1 kHz 0.29885 V 0.30115 V
300 mV 300 mV@100 kHz 0.2845 V 0.3155 V
3 V 3 V@1 kHz 2.9885 V 3.0115 V
30 V 30 V@1 kHz 29.885 V 30.115 V
300 V 300 V@1 kHz 298.85 V 301.15 V
Frequency Accuracy Test
The following procedure tests the accuracy of the frequency function for the PAI and FAI
modules. Measurement accuracy applies to all channels, not just the channel used for the
test.
1. Configure Channel 1 for Frequency Using the Fluke DAQ software, configure
channel 1 for frequency. There is no range selection for frequency as all frequency
measurements use auto-ranging.
2. Open Spy Open the Spy window under the Configuration dialog.
3. Verify Accuracy Configure the 5700A for the output values below and verify the
measurement is between the minimum and maximum values. Change the channel 1
range as required (see Step 1).
Frequency Range 5700A Output Minimum Reading Maximum Reading
Autorange Only 1 V@10 kHz 9.994 kHz 10.006 kHz