User's Manual

268XA
Service Manual
2-38
Table 2-13. Stallion Switch Settings (cont)
Function
Stallion Settings
S Switches Other Switches
Zero BR1 23 35 37 39 44 64 BR1 ACR4 ACR5 FPWR RPCTL
Zero BR2 23 35 37 39 44 64 BR2 ACR4 ACR5 FPWR RPCTL
Zero BR3 23 35 37 39 44 64 BR3 ACR4 ACR5 FPWR RPCTL
Zero BR4 23 35 37 39 44 64 BR4 ACR4 ACR5 FPWR RPCTL
REFBAL2 26 34 37 39 44 64 BR4 ACR4 FPWR RPCTL
REFBAL0 26 34 37 39 44 64 BR4 ACR4 FPWR RPCTL
REFJUNC 17 21 34 37 39 44 64 BR2 ACR4 FPWR RPCTL
OTC Dischg 17233437424344466465 BR1 ACR4 FPWR RPCTL
After the input channel has been selected and the Stallion chip programmed
appropriately, there is a minimum time required for the signal conditioning circuitry to
settle. This settling time varies depending on the function and range being measured, and
is given in Table 2-14.
Table 2-14. Signal Conditioning Settling Time
Function Time
V dc 30 µs
V ac, fast 100 ms
V ac, medium 150 ms
V ac, slow 200 ms
300 20 µs
3 k 100 µs
30 k 400 µs
300 k 2 ms
3 M 10 ms
Frequency, fast 100 ms
Frequency, medium 150 ms
Frequency, slow 200 µs
Zero, BR1 30 µs
Zero, BR2 30 µs
Zero, BR3 30 µs
Zero, BR4 30 µs
Reference Balance, both references on 30 µs
Reference Balance, both references off 30 µs
Reference Junction 30 µs