User's Manual

28 II EX Calibration Manual Supplement
11/12 1
Change #1, 62942
On page 8, replace Duty Cycle (Vdc and mVdc) with:
Duty Cycle (Vdc and mVdc)
Range Accuracy
0.0 % to 99.9 %
[1]
Within (0.2 % per kHz + 0.1 %) for rise times <1 s.
[2]
[1] 0.5 Hz to 200 kHz, pulse width >2 s. Pulse width range is determined by the frequency by the frequency of the signal.
[2] For 6 V dc range accuracy is unspecified.
Under MIN MAX Recording change:
From:
250 s (peak)
[1]
Specified accuracy 100 counts for changes >250 s in duration
(add 100 counts for readings over 6000 counts)
(add 100 counts for readings in Low Pass mode)
To:
250 s (peak)
[1]
Specified accuracy 200 counts for changes >250 s in duration
(add 100 counts for readings over 6000 counts)
(add 100 counts for readings in Low Pass mode)
On page 14, Step 19 replace the Range:
From: 6 V
To: 60 V
On page 15, Step 44 replace the UUT Measurements:
From
44 0.28 nf 0.48 nf
To:
44 0.70 nf 1.10 nf
Change #2
On page 14, Steps 14, 15 ,16 , 17 & 18 replace the Low Limit and the High Limit with:
Low Limit High Limit
99.93 kHz 99.97 kHz
199.48 kHz 199.52 kHz
99.93 kHz 99.97 kHz
99.93 kHz 99.97 kHz
999.8 Hz 1000.2 Hz
On page 15, Step 46, replace Low Limit and the High Limit with:
Low Limit High Limit
9.2 F 9.8 F