User's Manual
28 II EX Getting Started Manual Supplement
2 2/12
Ex-Certification Data
Ex-Type certificate no: PTB 11 ATEX 2028 X
IECEx PTB 11.0080X
Ex-Designation: ATEX: II 2G Ex ia IIC T4 Gb
II 2D Ex ia IIIC T130 C Db
I M1 Ex ia I Ma
IECEx: Ex ia IIC T4 Gb
Ex ia IIIC T130 C Db
Ex ia I Ma
CE: CE0102
Operating Temperature: -15 C to 50 C
Storage Temperature: -40 C to +60 C
Batteries: 3 AAA Alkaline batteries, NEDA 24A IEC
LR03. Table 7 shows the approved batteries for this
Product.
For connections to intrinsically-safe circuits, observe
these Product connections:
Voltage – measurement input U
i
= 65 V:
U
0
= 9.54 V
C
i
= negligible
I
0
= 3.7 mA I
i
= negligible
L
i
= negligible
P
0
= negligible Ri = 2.47K
Lo/Co
Lo/mH 1000 100 2 0.5 0.1 0.01
Co/F 0 0.61 1 1.4 2.1 3.6
Current – measurement input I
i
= 5 A:
U
0
= 0 V U
i
= 65 V
C
0
= 1000 F C
i
= negligible
I
0
= 0mA
L
0
= 1000 mH L
i
= negligible
P
0
= 0 mW
mA/A Jack
U
0
= 1.95 V U
i
= 65 V
C
i
= negligible
I
0
= 9.7 A I
i
= Internally limited by a 440 mA fuse
L
i
= negligible
P
0
= negligible