User's Manual
28 II EX Getting Started Manual Supplement
4/12 7
On page 25, under Temperature replace Storage with:
Storage -40 C to +85 C (without battery)
-40 C to +60 C (with battery)
Change #2, 62942
On page 32, replace Duty Cycle (Vdc and mVdc) with:
Duty Cycle (Vdc and mVdc)
Range Accuracy
0.0 % to 99.9 %
[1]
Within (0.2 % per kHz + 0.1 %) for rise times <1 s.
[2]
[1] 0.5 Hz to 200 kHz, pulse width >2 s. Pulse width range is determined by the frequency by the
frequency of the signal.
[2] For 6 V dc range accuracy is unspecified.
On page 33, under MIN MAX Recording change:
From:
250 s (peak)
[1]
Specified accuracy 100 counts for changes >250 s in duration
(add 100 counts for readings over 6000 counts)
(add 100 counts for readings in Low Pass mode)
To:
250 s (peak)
[1]
Specified accuracy 200 counts for changes >250 s in duration
(add 100 counts for readings over 6000 counts)
(add 100 counts for readings in Low Pass mode)