User's Manual

Manual Supplement 28 II EX Getting Started
8 6/13
Change #3, 65782
On page 2, under Safety Information remove The Product complies with: and all the bullets.
On page 8, in the Symbols table, remove CAT III, CAT IV and CSA C1 and add:
Conforms to relevant South Korean EMC Standards
CAT II
Measurement Category II is applicable to test and measuring circuits connected directly
to utilization points (socket outlets and similar points) of the low-voltage MAINS
installation.
CAT III
Measurement Category III is applicable to test and measuring circuits connected to the
distribution part of the building’s low-voltage MAINS installation.
CAT IV
Measurement Category IV is applicable to test and measuring circuits connected at the
source of the building’s low-voltage MAINS installation.
On page 26, replace Electromagnetic Compatibility and IP Rating with the following, and
remove Shock, Safety Compliance and Certifications and add:
IP Rating .......................................................... IEC 60529: IP67 (Non-operating)
Safety ............................................................... IEC 61010-1: 600 V CAT IV / 1000 V CAT III, Pollution Degree 2
Electromagnetic Environment ....................... IEC 61326-1: Portable
Electromagnetic Compatibility ...................... In an RF field of 3 V/M, accuracy = specified accuracy +20 counts, except 600 A
dc range total accuracy = specified accuracy +60 counts. Temperature not
specified
Applies to use in Korea only. Class A Equipment (Industrial Broadcasting &
Communication Equipment)
[1]
[1] This product meets requirements for industrial (Class A) electromagnetic
wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and is not to be