Product Manual

Earth Ground Tester
Specifications
17
Measuring
Range
Resolution Display Range
Intrinsic
Uncertainty
Operating
Uncertainty
IEC 61557*
0.15 Ω to 20 Ω 0.01 Ω 0 Ω to 19.99 Ω
±(6 % of
measured
value + 5D)
±(18 % of
measured
value + 5D)
200 Ω 0.1 Ω 20 Ω to 199.9 Ω
2 kΩ 1 Ω 200 Ω to 1999 Ω
*Covers all deviations caused by influence quantities E
1
-E
5
If the deviation E4 caused by high probe or auxiliary probe resistance is higher
than specified W flashes. Measured values are outside of the specified
operating uncertainty.
Measuring time: 8 seconds (average from when START is pressed)
Limit input: Tester retains set value even if instrument is turned off
(assuming battery power supply is sufficient)
If Tester detects stray interference voltage 20 V, W is displayed and the
measurement is not started.