Manual Supplement Manual Title: Part Number: Print Date: Revision/Date: 287/289 Users Web-Only June 2007 2, 3/09 Supplement Issue: Issue Date: Page Count: 6 10/13 5 This supplement contains information necessary to ensure the accuracy of the above manual. © 2010-2013 Fluke Corporation. All rights reserved.
Manual Supplement 287/289 Users Change #1, 55189, 55213, 55269 On page 72, under Resistance Specifications, change the Accuracy for 500 k: From: 0.05 % + 2 To: 0.05 % + 15 On page 74, under Frequency Counter Specifications, add a footnote 3 to the Duty Cycle row. [3] For 10 s < pulse width <25 s add 1%. For 2 s < pulse width 10 s add 3.5 %. Under Pulse Width, change the Accuracy for 0.1000 ms: From: 0.002 ms + 3 counts To: 0.
Manual Supplement 287/289 Users Change #2 On page 63, Table 11, replace Items 23 and 24 with the following: 23 TL175 Test Lead Set 1 3504931 24 Alligator Clips, one black and one red 2 3970052 (Black) 3970041 (Red) On page 65, Table 12, replace the first Item with the following AC172 Alligator Clips for use with the TL175 test lead set Change #3, 54932, 64414 On page 1, delete the Safety Information bullet section.
Manual Supplement 287/289 Users On page 4, replace the Symbols Table with: Symbol Description Symbol Description AC (Alternating Current or Voltage) Fuse DC (Direct Current or Voltage) Double Insulated Hazardous voltage Battery (Low battery when shown on the display) Continuity test or continuity beeper tone Conforms to relevant Canadian and US standards Conforms to European Union directives Conforms to relevant Australian standards Underwriters Laboratory
Manual Supplement 287/289 Users On page 64, replace Figure 31 with: 1 23 2 14 3 16 4 5 24 15 18 25 17 7 6 4 PL 19 8 2 PL 20 ed Trust rld. o Most The in the W Tools 9 10 21 22 11 6 PL 12 om Go to luke.c uct, d w.f your prodan d fin wwgis ter nuals tion to re load ma orma inf down product more 13 est40.eps Figure 31.
Manual Supplement 287/289 Users On page 63, Table 11, replace item 25, delete item 26 and replace the note with: 25 Fluke 287/289 Information Pack 4277223 To ensure safety, use the exact replacement only. [1] The Users manual is available through www.Fluke.com. Click on Support and then Product Manuals. On page 66 & 67 delete the Vibration, Shock, Safety Standard, Electromagnetic Compatibility Standards (EMC) and the Certifications sections and replace with: Safety ..................................