User's Manual

a3000 FC
Calibration Manual
6
Required Equipment
The equipment in Table 2 is necessary for performance tests and calibration
adjustment.
Table 2. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator
4.5-digit resolution
AC Current Accuracy:
600 μA to 10 A ±0.25 %
Fluke 5522A Calibrator
(or equivalent)
Wired coil 50 turns 5500A/COIL
Performance Tests
Warning
To prevent possible electrical shock, fire, or personal injury,
do not perform the performance test procedures unless the
Product is fully assembled.
The performance tests verify the full operation of the Product and measure the
accuracy of each function against Product specifications. If the Product fails a
part of the test, calibration adjustment and/or repair is necessary. See
“Calibration Adjustment”.
Test the Display
To verify that all segments of the display function:
1. With the Product OFF, push and hold .
2. Push while you keep pushed until all of the display segments are
shown. See Figure 2.
hby002.eps
Figure 2. All Segments of the Display
If segments of the display are missing, repair is necessary. See “Contact
Fluke”.