Fluke 750 Series Documenting Process Calibrators: Work smarter. Work faster.
Fluke 750 Series Documenting Process Calibrators: Work Smarter. Work Faster. Whether you’re calibrating instruments, troubleshooting a problem or running routine maintenance, Fluke 750 Series process calibrators can help you get the job done faster. It does so many different tasks, so quickly and so well, it’s the only process calibrator you need to carry. • Multi-lingual interface displays instructions in English, French, German, Spanish, and Italian. • Multifunctional.
Fluke 750 Series Documenting Process Calibrators: Calibrators as versatile as you are. The Fluke 750 Calibrators, offered in two models, let you choose the right set of capabilities for your needs. Capability 753 754 Source /measure • • • The Fluke 753 offers simultaneous source and measure capabilities for all common process parameters. Create and execute automated procedures and automatically capture the results results.
It’s easy to calibrate and maintain HART instrumentation with one powerful tool. + – With the 754 DPC, you can: • Generate precision electrical, temperature, or pressure signals for analog stimulus or sensor simulation. • Simultaneously measure electrical, temperature, or pressure signals from transmitter output. • Re-label smart transmitters by reading and writing HART tag and message fields. • Clone additional transmitters by reading and storing basic HART configurations.
HART applications Fluke 754: The HART calibrator that is easy to use. Versatile HART protocol support The 754 supports the commands contained in HART protocol version 5.7.
Why use “smart” instrumentation? Like most process plants, your organization is probably facing the dual challenges of maximizing productivity while minimizing maintenance costs. “Smart” digital transmitters offer superior performance and reliability, while saving time and effort in maintenance and calibration. Manufacturers of field instruments have helped accelerate the changeover by offering smart transmitters at prices nearly as low as analog units.
Automatically record and document your work Fluke 750SW DPC/TRACK 2 software includes aninstrumentation data-base that makes it easy to manage your instrumentation, create and schedule tests, load and unload the 753 or 754, print a variety of standard reports, and manage calibration data. Print standard reports automatically. The software assembles pre-formatted reports from your database files, saving time and reducing errors.
Measurement function specifications Confidence Interval: k=3 DC voltage measurement Range (full scale) 100.000 mV 3.00000 V 30.0000 V 300.00 V Accuracy (% of reading + floor) 1 year 2 years 0.02 % + 0.005 mV 0.03 % + 0.005 mV 0.02 % + 0.00005 V 0.03 % + 0.00005 V 0.02 % + 0.0005 V 0.03 % + 0.0005 V 0.05 % + 0.05 V 0.07 % + 0.05 V Temperature coefficient: (0.001 % reading + 0.0002 % range)/°C from -10 °C to 18 °C and 28 °C to 50 °C, 100.000 mV range: 0.001 % of reading + 0.
Temperature measurement and simulation specifications Confidence Interval: k=3 Temperature, Resistance Temperature Detectors1 Degrees or % of reading Type (α) Range °C 100 Ω Pt (385) 200 Ω Pt (385) 500 Ω Pt (385) 1000 Ω Pt (385) 100 Ω Pt (3916) 100 Ω Pt (3926) 10 Ω Cu (427) 120 Ω Ni (672) -200 to 100 100 to 800 -200 to 100 100 to 630 -200 to 100 100 to 630 -200 to 100 100 to 630 -200 to 100 100 to 630 -200 to 100 100 to 630 -100 to 260 -80 to 260 Measure °C2 1 year 2 years 0.07 °C 0.14 °C 0.02 % + 0.
Pressure specifications The Fluke family of 50 pressure modules: Covers virtually any pressure application including gage, differential, dual (compound), absolute, and vacuum. • Display pressure readings in any of ten different pressure units you specify in the calibrator setup.
Pressure module specifications (cont.) (all specifications in % of full span. Specifications reflect a confidence interval of 95 %.) Parameter/Range Hi Side Media2 Lo Side Media2 Reference Total Uncertainty Uncertainty4 1-year (15-35 �C) Total Uncertainty 1-year1 Total Uncertainty 6-month (15-35 �C) 750PV3 -5 psi (-350 mBar) 316 SS Dry Air ± 0.03 % ± 0.06 % ± 0.07 % ± 0.05 % 750PV4 -15 psi (-1 bar) 316 SS Dry Air ± 0.03 % ± 0.06 % ± 0.07 % ± 0.
General specifications Ordering information Data log function Measure functions: Voltage, current, resistance, frequency, temperature, pressure Reading rate: 1, 2, 5, 10, 20, 30, or 60 readings per minute Maximum record length: 8000 readings (7980 for 30 or 60 readings per minute) FLUKE-753 Documenting Process Calibrator FLUKE-754 Documenting Process Calibrator-HART Standard accessories include: Three sets of stackable test leads, three sets of TP220 test probes with three sets of “extended tooth” alliga