Product Manual

Documenting Process Calibrator
Functions
9
Functions
A summary of functions supplied by the Product is shown in
Table 2. More features include:
Analog display for easy to read measurements when
inputs are unstable.
Localized display (5 languages). See “Display
Languages”.
Thermocouple (TC) input/output jack and internal
isothermal block with automatic reference-junction
temperature compensation. Or manually record an
external temperature reference.
Test results storage.
Data logging. Automatically log up to 8,000 data points.
A USB computer interface to upload or download tasks,
lists, and results.
Automatic calibration procedures for transmitters and
limit switches when you use split screen
MEASURE/SOURCE mode.
Transmitter mode in which the Product can be
configured to emulate the functions of a process
instrument.
Calculator feature with square-root function, and
accessible registers that contain measure and source
values.
Damp feature (smoothes the last several readings),
with display indicator of damped status.
Display of measurements in engineering units, percent
of scale, square-law inputs, or custom units.
Min/Max feature captures and shows minimum and
maximum measured levels.
Set source values to engineering units, percent of
scale, square-law outputs, or custom units.
Manual and automatic stepping and an output ramp
feature for testing limit switches. Trip detect is either a 1
V change or a continuity status change (Open or Short)
from one ramp increment to the next.
For performance testing and calibration instructions,
download the 753/754 Calibration Manual from Fluke’s
website.