Product Manual
Documenting Process Calibrator
The Calculator
87
Run a Preloaded Task
Push the More Choices softkey until the Tasks softkey is
shown, then push Tasks to see the list of tasks
(procedures) downloaded from a host computer. Tasks are
Product configurations, kept with a procedure name, for
example the type and manufacturer of a specific transmitter.
A task configures the Product for transmitter calibration with
all the calibration parameters (source and measure
functions, 0% and 100% levels, test strategy) predefined.
While the task controls the Product, the Continue softkey
becomes Continue Task.
Clear the Memory
In Setup mode, highlight the Clear Memory choice and
push to erase the memory:
• Results that have been kept
• Min Max data
• Log data sets
A confirmation message is shown so that the memory is not
accidently erased.
The Calculator
For mathematical equations that involve the Product’s
source or measured value, use the Product’s built-in
calculator. The current measure and source values and
units, are always available to be put into an equation with
one keystroke. The Product measures and sources during
calculator operation.
Push the Calc softkey to start the calculator from the
SOURCE, MEASURE, or MEASURE/SOURCE mode. Push
the More Choices softkey to see the Calc softkey if
necessary.
After you push Calc, the display, number keys, and keys
with calculator functions (, , , , r and )
become an algebraic-entry calculator.
Push the Done softkey to start normal Product operation.