700G Series Pressure Gauge Users Manual November 2011 Rev. 1, 10/13 © 2011-2013 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf.
Table of Contents Title Page Introduction .................................................................................................................... How to Contact Fluke ..................................................................................................... Standard Equipment....................................................................................................... Safety Information ......................................................................................
700G Series Users Manual Set Sample Rate .................................................................................................. Set Tare ............................................................................................................... Function Lock ....................................................................................................... Supervisory Mode .....................................................................................................
List of Tables Table 1. 2. Title Page Symbols................................................................................................................................. 4 Display and Buttons ..............................................................................................................
700G Series Users Manual iv
List of Figures Figure 1. 2. Title Page The Product ........................................................................................................................... 5 Change the Batteries .............................................................................................................
700G Series Users Manual vi
Introduction How to Contact Fluke The 700G Series Pressure Gauges (the Product) are high-accuracy digital pressure test gauges. Accurate to 0.05 % FS, the Products can be used as a calibration reference, or in any application where high-accuracy pressure measurement is required.
700G Series Users Manual Standard Equipment • Replace the batteries when the low battery indicator () shows to prevent incorrect measurements. • Do not use and disable the Product if it is damaged. • Read all safety Information before you use the Product. • Do not use the Product in damp or wet environments.
Pressure Gauge Safety Information Hazard Location Information/Approvals Ex Hazardous Areas An Ex-hazardous area as used in this manual refers to an area made hazardous by the potential presence of flammable or explosive vapors. These areas are also referred to as hazardous locations, see NFPA 70 Article 500. ® LR110460 Class I, Div. 2, Groups A-D II 3 G Ex nA IIB T6 KEMA 06ATEX0014 X Ta=–10 °C...
700G Series Users Manual Symbols Symbols used on the Product and in this manual are explained in Table 1. Table 1. Symbols 4 Symbol Meaning Symbol Meaning Risk of danger. Important information. See manual. Conforms to European Union directives. Hazardous voltage. Risk of electrical shock. Conforms to relevant North American Safety Standards. Pressure Conforms to relevant South Korean EMC Standards. Conforms to relevant Australian standards.
Pressure Gauge Display and Buttons Display and Buttons The Display and Buttons are shown in Figure 1. The Buttons are explained in Table 2. 2 3 700G31 10K PSIG 1 PRESSURE GAUGE 4 10K PSI 700 BAR gsn001.eps Figure 1.
700G Series Users Manual Table 2. Display and Buttons Item Function Push to turn the Product on. Push again to turn it off. Zeros the display. In Configure Mode, push the button to move forward through the menus. 6 Note For absolute versions of the gauge, pushing prompts you to enter a barometric reference pressure. Use and to adjust the reading as necessary and then push .
Pressure Gauge Table 2. Display and Buttons (cont.) Item Item Function Function ¼ inch NPT Connector MIN MAX records minimum and maximum pressure values and saves them in memory. Push to show maximum (MAX) indication. Push again to show minimum (MIN) indication. After 2 seconds, the gauge goes back to live operation. Pressure Display Engineering Units Bargraph To clear the MIN MAX memory values, push and hold for 2 seconds until CLr is shown.
700G Series Users Manual Operation The subsequent sections tell you how to operate the Product. Push to turn on the Product. The analog bar graph at the bottom of the display shows the applied-pressure level relative to the full range of the gauge. Note If you record a Tare value, the pressure shown is not the actual pressure applied. How to Setup the Product Before you use the product, it is necessary to configure it for your application. Push to go to the Setup menu. engineering units.
Pressure Gauge Operation Set Sample Rate Function Lock This function finds how often pressure is sampled and the display is updated. Selections are 0.5, 1, 3, and 10 samples/second. Note that 10/second gives the fastest response time. When set, access to each of the settable parameters above can be turned “off” to prevent unauthorized configuration changes. This is done with password protection in Supervisory mode. Push to access Supervisory mode or to go back to normal operation.
700G Series Users Manual each parameter. To move to the next parameter, push .
Pressure Gauge Maintenance How to Change the Batteries Maintenance How to Clean the Product Warning To prevent possible electrical shock, fire, or personal injury, batteries must only be changed in an area known to be nonhazardous. Explosion hazard. Clean the Product with a soft cloth dampened with water or water and weak soap. Caution To prevent possible damage to the Product, do not use solvents or abrasive cleansers. To change the batteries, see Figure 2: 1.
700G Series Users Manual Accessories RS-232 Interface The Product includes an RS-232 interface. Remove the Product holster and the input jack is on the back of the Product. You can use serial communication to configure and calibrate the Product and move measurement data from the Product to a PC. An RS-232/USB cable is sold separately and includes 700G/TRACK Software. For specifications on the interface, see the Specifications section.
Pressure Gauge Specifications Specifications Available Input Ranges See Ranges and Resolution table for available ranges in psi plus equivalent ranges and resolution for all engineering units. Accuracy 700G Ranges Positive Pressure ............................................................................ ±0.05 % FS Positive Pressure (700G01, 700G02).............................................. ±0.1 % FS Vacuum ...........................................................................................
700G Series Users Manual Media Compatibility 700G01, 700G02, 700G04, 700G05, 700RG05 ............................. any clean dry non-corrosive gas All other ranges, 15 psi to 1000 psi................................................. any liquids or gases compatible with 316 stainless steel Above 1000 psi ...............................................................................
Pressure Gauge Specifications Mechanical Specifications Dimensions .................................................................................... 11.4 x 12.7 (cm), depth = 3.7 cm (4.5 x 5 (in), depth= 1.5 in) Pressure Connection ............................................................................. ¼ in NPT Male Housing .................................................................................. Cast ZNAL Display 5-1/2 Digits, 16.53 mm (0.
700G Series Users Manual Ranges and Resolution Model Number Pressure Range (psi) Vacuum Range (psi) Burst Pressure (psi) Proof Pressure (psi) 700G01 700G02 700G04 700G05 700G06 700G27 700G07 700G08 700G10 700G29 700G30 700G31 0.4 1 15 30 100 300 500 1000 2000 3000 5000 10000 -0.
Pressure Gauge Specifications 700G01 700G02 700G04 700G05 700G06 700G27 700G07 700G08 700G10 700G29 700G30 700G31 Pressure Range (psi) Model Number 0.4 1 15 30 100 300 500 1000 2000 3000 5000 10000 Vacuum Range (psi) -0.
700G Series Users Manual Model Number 700GA4 700GA5 700GA6 700GA27 700RG05 700RG06 700RG07 700RG08 700RG29 700RG30 700RG31 Pressure Range (psi) 15 PSIA 30 PSIA 100 PSIA 300 PSIA 30 100 500 1000 3000 5000 10000 Vacuum Range (psi) 0 PSIA 0 PSIA 0 PSIA 0 PSIA -14 -12 -12 -14 -14 -14 -14 Burst Pressure (psi) 60 120 400 1200 90 400 2000 4000 10000 15000 20000 Proof Pressure (psi) 30 60 200 600 60 200 1000 2000 6000 10000 15000 Engineering Unit Factor psi
Pressure Gauge Specifications Model Number 700GA4 700GA5 700GA6 700GA27 700RG05 700RG06 700RG07 700RG08 700RG29 700RG30 700RG31 Pressure Range (psi) 15 PSIA 30 PSIA 100 PSIA 300 PSIA 30 100 500 1000 3000 5000 10000 Vacuum Range (psi) 0 PSIA 0 PSIA 0 PSIA 0 PSIA -14 -12 -12 -14 -14 -14 -14 Burst Pressure (psi) 60 120 400 1200 90 400 2000 4000 10000 15000 20000 Proof Pressure (psi) 30 60 200 600 60 200 1000 2000 6000 10000 15000 Engineering Unit Factor m
700G Series Users Manual 20