3 Fluke Calibration P3000 Series Hydraulic Deadweight Testers Models P3100 and P3200
Pressure ranges
Oil operated Ranges to 20,000 psi (1,400 bar)
Distilled water operated Ranges to 10,000 psi (700 bar)
Accuracy ± 0.015 % of reading (± 0.008 % optional)
Note: Accuracy based on % of reading from 10 % to
100 % of the piston range when used in accordance with
the corrections found on the calibration certificate. Below
10 % ± (accuracy class) x 10 % of the piston range.
Materials of construction
Standard weight material Series 3 non-magnetic austenitic stainless steel
Density: 7.8 g/cm
3
Optional fractional weights Solution heat treated aluminum
Density: 2.7 g/cm
3
Piston material Tungsten carbide with nickel binder
Density: 15.0 g/cm
3
Cylinder material Hardened martensitic steel
Tungsten carbide (water above 500 psi, 35 bar)
Thermal coefficients of expansion Oil piston/cylinder 16.5 ppm/°C
Water (above 500 psi, 35 bar) 11 ppm/°C
General
Test port adaptors M14 X 1.5, M20 X 1.5, 1/8 in, 1/4 in, 3/8 in and 1/2 in
NPT and BSP
Weight 16 kg (36 lb)
Dimensions (W x D x H) 440 mm x 300 mm x 215 mm (17.5 in x 12 in x 8.5 in)
Mass set weight (typical) 36 kg (80 lb)
Reservoir volume 150 cm
3
(9.2 in
3
)
Screw press displacement 5.5 cm
3
(0.34 in
3
)
Pump displacement 4.7 cm
3
(0.29 in
3
) per stroke
O-ring seal materials Viton
®
Weight increments
Minimum standard weight increments
Up to 500 psi (35 bar) 1 psi (0.1 bar)
500 psi to 5 000 psi (35 bar to 350 bar) 10 psi (1 bar)
5 000 psi to 20 000 psi (350 bar to 1 400 bar) 20 psi (2 bar)
Model P3112 1 psi (0.1 bar)
Optional fractional weights
Up to 500 psi (35 bar) 0.1 psi (0.01 bar)
500 psi to 5 000 psi (35 bar to 350 bar) 1 psi (0.1 bar)
5 000 psi to 20 000 psi (350 bar to 1400 bar) 2 psi (0.2 bar)
Operating fluids
Mineral oil Shell spindle oil 22, our reference 55-655, viscosity 40 cs@ 20 °C (68 °F)
Distilled or deionized water
Specications