Product Manual
3-1
Chapter 3
Priming
Priming
1. Open reservoir valve one turn counter-clockwise and turn capstan fully in.
2. Pump the priming pump two times.
3. Close valve and turn capstan fully out.
4. Open valve and turn capstan fully in.
Note
During this operation, bubbles may appear in the reservoir, as trapped air
is expelled. For large volumes, repeat steps 3 and 4 until no further bubbles
appear.
5. With valve open, turn capstan fully out and close valve. The tester is now ready for
use.
W Warning
Turning the capstan out with the reservoir valve closed will
generate approximately 15 inHg / 0.5 bar vacuum. If the DUT is
vacuum sensitive, leave valve open during priming operation.