Item Brochure

2 Fluke Calibration P3800 Series High Pressure Hydraulic Deadweight Testers
Pressure ranges
P3830 500 psi to 30 000 psi, or 40 bar to 2 000 bar, or 4 MPa to 200 MPa
P3840 500 psi to 40 000 psi, or 40 bar to 2 600 bar, or 4 MPa to 260 MPa
P3860 500 psi to 60 000 psi or 40 bar to 4 000 bar or 4 MPa to 400 MPa
Accuracy Standard accuracy is 0.02 % of Reading*. Optional accuracy of 0.015 % of Reading is available. ISO/IEC
17025 accredited calibration certificate is standard
*Accuracy is based on % of Reading from 10 % to 100 % of the piston range when used in accordance with
the corrections found on the calibration certificate. Below 10 %, ± (accuracy class) x 10 % of the piston range.
Materials of construction
Standard weight material Series 3 non-magnetic, austenitic, stainless steel
Density: 7.8 g/cP
3
Piston material Tungsten carbide with nickel binder
Cylinder material Tungsten carbide with cobalt binder
Thermal coefficient of
expansion
11 ppm/°C
General
Test port adaptors 3/4-16 UNF; 9/16 in – 18 UNF (Autoclave); 3/8 in BSP; 1/2 in BSP and 3/4 in BSP
Weight 30 Kg (66 lb) Instrument base only
Dimensions (W x D x H) 455 mm X 340 mm X 478 mm ( 18 in X 13.5 in X 19 in)
Reservoir volume 235 cm
3
(14.3 in
3
)
Intensifier ratio 7 to 1
Seal materials Nitrile
Operating fluid Sebacate, our reference 3069551
Specications
Ordering information
Models:
P3830-BAR High Pressure Hydraulic Deadweight Tester, 40 to 2 000 bar
P3830-PSI High Pressure Hydraulic Deadweight Tester, 500 to 30 000 psi
P3830-MPA High Pressure Hydraulic Deadweight Tester, 4 to 200 MPa
P3840-BAR High Pressure Hydraulic Deadweight Tester, 40 to 2 600 bar
P3840-PSI High Pressure Hydraulic Deadweight Tester, 500 to 40 000 psi
P3840-MPA High Pressure Hydraulic Deadweight Tester, 4 to 260 MPa
P3860-BAR High Pressure Hydraulic Deadweight Tester, 40 to 4 000 bar
P3860-PSI High Pressure Hydraulic Deadweight Tester, 500 to 60 000 psi
P3860-MPA High Pressure Hydraulic Deadweight Tester, 4 to 400 MPa
Options and accessories
A. Improved Accuracy Option with Software - PressCal Software is a
Windows based software program that allows users to easily apply all
necessary corrections, allowing for improved accuracy of 0.015%.
B. Fine Increment Weights - Additional small weights that allow for finer
pressure increments
PPA9159-CAL For use with kPa, MPa, bar, or kgf/cm2 weight sets
PPA9608-CAL For use with psi weight sets
C. Conversion Weights - Allows for an instrument to be used in a pressure
unit other than the pressure unit the instrument was ordered for.
Fluke Calibration
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (877) 355-3225 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
©2011-2012 Fluke Calibration.
Specifications subject to change without notice.
Printed in U.S.A. 8/2012 3833630C D-EN-N
Modification of this document is not permitted
without written permission from Fluke Calibration.
Fluke Calibration.
Precision, performance, confidence.