Product Manual

Remote Commands
Remote Command List 6
6-11
*IDN?
This command shows the unit identification: instrument manufacturer, model number,
serial number, and firmware revision level.
Parameter: <None>
Response: <text string>
where <text string> contains the subsequent four fields
separated by commas:
1. Manufacturer
2. Model number
3. Serial number (always 0)
4. Firmware revision level
Example: *IDN?
FLUKE, Calibrator,0,1.2
This example is an indication the manufacturer is Fluke, the
model is Calibrator, the serial number is 0, and the firmware
version is 1.2.
ISO_MEAS
This command sets the isolated measurement type.
Parameter: <value>
where <value> is one of these:
DC10V measure DC voltage, 10V range
DC100V measure DC voltage, 100V range
DCI measure DC current
PRESSURE measure pressure
Response: <None>
Example: ISO_MEAS DCI
This example sets the isolated measurement to DC current.