Product Manual

7526A
Users Manual
6-14
LOCKOUT
This command puts the Calibrator into the lockout condition when in remote control (see
the REMOTE command). In this condition, no local operation is permitted at the front
panel. This includes the LOCAL key. To clear the lockout condition, use the LOCAL
command. This command duplicates the IEEE-488 LLO (Local Lockout) message.
Parameter: <None>
Response: <None>
Example: LOCKOUT
This example puts the instrument into the lockout condition.
No front panel controls can be used.
LOOP_POWER?
This command shows the isolated DC current input 24V loop power status.
Parameter: <None>
Response: <value>
where <value> is ON or OFF
Example: LOOP_POWER?
OFF
This example is an indication that the isolated DC current
input 24V loop power is turned off.
LOOP_POWER_OFF
This command disables the isolated DC current input 24V loop power.
Parameter: <None>
Response: <None>
Example: LOOP_POWER_OFF
This example disables the isolated DC current input 24V loop
power.
LOOP_POWER_ON
This command enables the isolated DC current input 24V loop power.
Parameter: <None>
Response: <None>
Example: LOOP_POWER_ON
This example enables the isolated DC current input 24V loop
power.