Remote Commands
Remote Command List 6
6-15
*OPC
Operations Complete command. This command sets bit 0 (OPC) of the Event Status
Register to 1 when all pending device operations are complete. Also see the *ESR?
command.
Parameter: <None>
Response: <None>
Example: *OPC
This example sets bit 0 of the Event Status Register to 1 when
all pending device operations are done.
*OPC?
Operations Complete query. This command shows a 1 after all pending Calibrator
operations are complete. This command does not answer until all pending Calibrator
operations are complete, and causes the control program execution to pause until
operations are complete. Also see the *WAI command.
Parameter: <None>
Response: 1
Example: *OPC?
1
This example does not answer until all pending Calibrator
operations are complete, and then shows 1.
OPER
This command puts the Calibrator in operate mode and start the output at the front panel
terminals. This command does the same as if you push the front panel key when in
standby mode.
Parameter: <None>
Response: <None>
Example: OPER
This example connects the selected output to the Calibrator
front panel terminals. It also shows Opr on the display.