Product Manual

7526A
Users Manual
6-16
OPER?
This command shows the current operate/standby mode setting.
Parameter: <None>
Response: <value>
where <value> is "1" for operate mode and "0" for standby
mode
Example: OPER?
1
This example is an indication that the Calibrator is in operate
mode.
*OPT?
This command shows a list of the installed hardware and software options. This
command is reserved for future use.
Parameter: <None>
Response: <text string>
where <text string> is "0" if no options are installed, or a list
of installed options separated by commas
Example: *OPT?
0