Product Manual

Remote Commands
Remote Command List 6
6-17
OUT
This command sets the output mode and value of the Calibrator. To source a temperature,
select the necessary mode and sensor parameters first with the TSENS_TYPE,
RTD_TYPE, and TC_TYPE commands. Use the multiplier prefixes k for kilo, m for
milli, and u for micro with the OUT command units, as necessary. The unit can be
ignored if the output mode is not to be changed.
Parameter: <value> <units>
where <value> is the output value to be made and where
<units> is one of the items below, with a possible prefix
multiplier (k, m, or u) as given above:
V DC volts
A DC current
OHM Resistance
CEL Temperature in Celsius
FAR Temperature in Fahrenheit
Response: <None>
Examples: Explanation
OUT 15.2 V Change to DC volts, output 15.2 V
OUT 1.2 mA Change to DC current, output 1.2 mA,
note units prefix multiplier m
OUT 5 Ohm Change to Resistance, output 5 O
OUT 100 CEL Change to temperature in °C, output
100 °C
OUT 3 No change to output mode, output a
value of 3 in the current units