Product Manual

7526A
Users Manual
6-18
OUT?
This command shows the current output value and units of the Calibrator.
Parameter: <None>
Response: <value>,<units>
where <value> is the current output value and where <units>
is one of these:
V DC volts
A DC current
OHM Resistance
CEL Temperature in Celsius
FAR Temperature in Fahrenheit
Examples: Explanation
OUT?
1.88300E-02,A Current output is 18.83 mA
OUT?
1.23000E+00,V Current output is 1.23 V
OUT?
4.00000E+03,OHM Current output is 4.0 KO
OUT?
1.0430E+02,CEL Current output is 104.3 °C
PRES?
This command queries the connected pressure module for its manufacturer, serial
number, and firmware version.
Parameter: <None>
Response: <text string>
where <text string> contains the subsequent three fields
isolated by commas:
1. Manufacturer
2. Serial number
3. Firmware revision level (always 0)
Example: PRES?
FLUKE,610070,0
This example is an indication that the manufacturer is Fluke,
the serial number is 610070, and the firmware version is 0.