Product Manual

Remote Commands
Remote Command List 6
6-21
RANGE?
This command shows the current DC voltage or current output range.
Parameter: <None>
Response: <value>
where <value> is one of these:
V_0.1V DC volts, 100 mV range
V_1V DC volts, 1V range
V_10V DC volts, 10 V range
V_100V DC volts, 100 V range
A_0.1A DC current
NONE neither volts nor current is selected
Example: RANGE?
V_10V
This example is an indication that the current output range is
10 V.
RANGELCK
This command locks or unlocks the DC voltage output range in reference to the current
output value.
Parameter: <value>
where <value> is one of these:
ON to lock the current voltage range
OFF to unlock the current voltage range
Response: <None>
Example: RANGELCK ON
If the current DC voltage output is 5 V, this example locks the
range at 10 VDC.