Product Manual

Remote Commands
Remote Command List 6
6-23
*RST
This command resets the Calibrator to the power-up condition and holds off execution of
subsequent commands until the reset operation is complete.
A reset action invokes the subsequent commands and values for the primary display:
Command Value
OUT 0 V
PRES_UNIT Last selected
RANGE 0.1 V
RTD_TYPE Last selected
STBY (No output)
TC_REF INT
TC_TYPE Last selected
TSENS_TYPE Last selected
The isolated display and selections stay as they were last selected.
Parameter: <None>
Response: <None>
Example: *RST
This example resets the Calibrator, invoking the commands
and values shown above.
RTD_MEAS
This command puts the primary display in RTD measure mode.
Parameter: <value>
where <value> is one of the these:
CEL display in degrees Celsius
FAR display in degrees Fahrenheit
<None> display in the last selected temperature
unit
Response: <None>
Example: RTD_MEAS CEL
This example puts the Calibrator to RTD measure mode and
shows in degrees Celsius.