STBY
This command puts the Calibrator in standby mode. This stops the output at front panel
terminals. This command does the same as if you push the front panel key when in
operate mode.
Parameter: <None>
Response: <None>
Example: STBY
This example disconnects the selected output from the
Calibrator front panel terminals. It also shows Stby on the
display.
ST_START
This command puts the Calibrator in switch test mode and starts a new switch test. If the
calibrartor is already is switchtest mode it simply starts a new switch test.
Parameter: <None>
Response: <None>
Example: ST_START
This example puts the calibrator in switch test mode and starts
a new switch test
ST_CLOSE?
In switch test recall mode this command shows the value at which the switch closed.
Parameter: <None>
Response: <value>,<units>
where <value> is the value recorded from the primary display
at the time the switch closed and where <units> are the units
recorded from the primary display at the time the test was
done.
Example: ST_CLOSE?
5.00000E+01,CEL
This example shows the value at which a temperature switch
closed in Celsius.