ST_OPEN?
In switch test recall mode, this command shows the value at which the switch opened.
Parameter: <None>
Response: <value>,<units>
where <value> is the value recorded from the primary display
at the time the switch opened and where <units> are the units
recorded from the primary display at the time the test was
done.
Example: ST_OPEN?
5.53000E+01,CEL
This example shows the value at which a temperature switch
opened in Celsius.
ST_DEAD?
In switch test recall mode this command shows the dead band of the switch.
Parameter: <None>
Response: <value>,<units>
where <value> is the deadband of the switch. To calculate
this, subtract the second recorded contact transition from the
first recorded contact transition and where <units> are the
units recorded from the primary display at the time the test
was done.
Example: ST_DEAD?
5.30000E+00,CEL
This example shows the value of the calculated deadband
temperature of the switch in Celsius.