Product Manual

Remote Commands
Remote Command List 6
6-29
TC_MEAS
This command places the primary display in thermocouple measure mode.
Parameter: <value>
where <value> is one of these:
CEL display in degrees Celsius
FAR display in degrees Fahrenheit
<None> display in the last selected temperature unit
Response: <None>
Example: TC_MEAS FAR
This example puts the Calibrator into thermocouple measure
mode, and shows in degrees Fahrenheit.
TC_REF
This command selects the internal temperature sensor or an external reference value for
cold junction compensation of thermocouple source and measurement.
Parameter: <value>
where <value> is one of these:
INT use internal temperature sensor
EXT use external reference value
Response: None
Example: TC_REF EXT
This example sets the thermocouple reference to external.
TC_REF?
This command shows the source of the temperature that is used for cold junction
compensation of thermocouple source and measurement.
Parameter: <None>
Response: <value>
where <value> is one of these:
INT internal temperature sensor in use
EXT external reference value in use
Example: TC_REF?
INT
This example is an indication that the internal sensor is in use.