Product Manual

Remote Commands
Remote Command List 6
6-31
TC_TYPE?
This command shows the Thermocouple (TC) sensor type that is used for TC temperature
source and measurement.
Parameter:
Response: <None>
<value>
where <value> is one of these:
B B-type thermocouple
C C-type thermocouple
E E-type thermocouple
J J-type thermocouple
K K-type thermocouple
L L-type thermocouple
N N-type thermocouple
R R-type thermocouple
S S-type thermocouple
T T-type thermocouple
U U-type thermocouple
XK XK-type thermocouple
BP BP-type thermocouple
Z 1 mV/°C
Example: TC_TYPE?
K
This example is an indication that the thermocouple sensor
type is K-type.
TSENS_TYPE
This command puts the temperature mode to thermocouple (TC) or Resistance
Temperature Detector (RTD) for temperature source and measurement.
Parameters: <value>
where <value> is one of these:
TC Thermocouple
RTD Resistance Temperature Detector
Response: <None>
Example: TSENS_TYPE RTD
This example sets the temperature mode to RTD.