Product Manual

7526A
Users Manual
6-32
TSENS_TYPE?
This command shows the current temperature mode, thermocouple (TC) or Resistance
Temperature Detector (RTD).
Parameter: <None>
Response: <value>
where <value> is one of these:
TC Thermocouple
RTD Resistance Temperature Detector
Example: TSENS_TYPE?
TC
This example is an indication that the current temperature
mode is thermocouple.
*TST?
This command does a series of self-tests and shows a "0" for pass or a "1" for fail. If
faults are found, they are put into the fault queue where they can be read by the FAULT?
query.
Parameter: <None>
Response: <value>
where <value> is one of these:
0 failed self-test
1 passed self-test
Example: *TST?
1
This example runs the self-test and shows that it passed.