Product Manual

Remote Commands
Remote Command List 6
6-33
VAL?
This command shows the last values for the isolated and primary measurements, in that
sequence.
Parameter: <None>
Response: <iso-value>,<iso-units>,<primary-value>,<primary-units>
where <iso-value> is the current isolated measurement
expressed in scientific notation.
and where <iso-units> is one of these:
V DC volts
A DC current
OVER measurement is over or under range
or one of the pressure units listed with the
ISO_PRES_UNIT? command
and where <primary-value> is the current primary
measurement expressed in scientific notation.
and where <primary-units> is one of these:
FAR °F
CEL °C
OHM Ohms
V DC volts (thermocouple millivolts)
OVER measurement is over or under range
NONE primary display is currently set to a
source mode
or one of the pressure units listed with the
PRES_UNIT? command
Example: VAL?
2.137000E+00,V,5.763300E+01,CEL
This example is an indication that the isolated measurement is
2.137 volts and that the primary measurement is 57.633 °C.