Product Manual

Introduction
Product Description 1
1-15
Figure 1-7 shows the primary thermocouple and RTD display.
1
2
3
654
gwp007.eps
Item Description
Output mode selection: RTD, TC, or rem for remote
operation
Input or output
selection
RTD or thermocouple type
selection
Output condition for RTD outputs:
Stby Standby, terminals inactive
Opr Operating, terminals are active with output per the shown value
Blank for RTD
inputs
Cold junction selection for thermocouple inputs and
outputs:
XCJC External cold junction compensation. The Product automatic cold
junction
compensation is turned off, for example, 0 mV is always
0 °C.
Blank Internal cold junction compensation. The Product automatically
measures the ambient temperature at the thermocouple terminals
and adjusts the measurement, for example, 0 mV is ambient
temperature.
Input or output value
Units
Figure 1-7. Primary Thermocouple and RTD Display