Product Manual

7526A
Users Manual
1-22
TC Type
Range (°C)
Absolute Uncertainty, tcal ±5 °C, ±(°C)
[1]
Output/Input
Minimum Maximum 90 days 1 Year
T
-250 °C
-200 °C
-100 °C
0 °C
200 °C
-200 °C
-100 °C
0 °C
200 °C
400 °C
0.34 °C
0.14 °C
0.09 °C
0.07 °C
0.06 °C
0.35 °C
0.16 °C
0.11 °C
0.09 °C
0.09 °C
U
-200 °C
0 °C
200 °C
0 °C
200 °C
600 °C
0.15 °C
0.08 °C
0.07 °C
0.16 °C
0.10 °C
0.10 °C
XK
-200 °C
-100 °C
0 °C
600 °C
-100 °C
0 °C
600 °C
800 °C
0.10 °C
0.07 °C
0.06 °C
0.07 °C
0.11 °C
0.09 °C
0.08 °C
0.09 °C
BP
0 °C
200 °C
600 °C
800 °C
1600 °C
2000 °C
200 °C
600 °C
800 °C
1600 °C
2000 °C
2500 °C
600
°
C
0.17 °C
0.14 °C
0.15 °C
0.22 °C
0.26 °C
0.38 °C
800
°
C
0.18°C
0.16 °C
0.17 °C
0.23 °C
0.28 °C
0.40 °C
[1] Does not include thermocouple wire error.
Type B, E, J ,K, N, R, S and T are based on ITS-90
Type L and U are based on DIN 43710-1985
Type C is based on ASTM standard E 988-96
Type XK and BP are based on GOST R 8.585-2001
RTD and Thermistor Specification, Output
RTD Types
Range °C
Absolute Uncertainty, tcal ±5 °C ±(°C)
[1]
Minimum 1 Year 90 Days 1 Year
Pt 385, 100 Ω -200 °C -800 °C 0.04 °C 0.05 °C
Pt 3926, 100 Ω -200 °C 630 °C 0.04 °C 0.05 °C
Pt 3916, 100 Ω -200 °C 630 °C 0.04 °C 0.05 °C
Pt 385, 200 Ω
-200 °C
400 °C
400 °C
630 °C
0.35 °C
0.42 °C
0.40 °C
0.50 °C
Pt 385, 500 Ω -200 °C 630 °C 0.15 °C 0.17 °C
Pt 385, 1000 Ω -200 °C 630 °C 0.07 °C 0.09 °C
Ni 120, 120 Ω -80 °C 260 °C 0.02 °C 0.02 °C
Cu 427, 10 Ω
[2]
-100 °C 260 °C 0.30 °C 0.38 °C
YSI 400
15 °C 50 °C 0.005 °C 0.007 °C
[1] 2-wire output
[2] Based on MINCO Application Aid No. 18