Product Manual

Primary Inputs and Outputs
Thermocouple (TC) Measure 2
2-9
custom SPRT, and push E .
7. At the "ENTER RTPW" prompt, key in the nominal resistance value (RTPW) for the
custom SPRT, and push E.
8. At the "ENTER COEFF A" prompt, key in the first (A) deviation coefficient for the
custom SPRT, and push E . To key in a coefficient that includes an exponent, key
in the mantissa, push E and . to select the EXP function, key in the exponent,
and push E.
9. When prompted, use the same method to key in the second (B), third (C), fourth
(A-), and fifth (B-) deviation coefficients.
10. To abort the SPRT entry without stored changes, push .
To use a custom SPRT:
1. Select RTD measure mode as given in the "Resistance Temperature Detector (RTD)
and Ohms Measure" section of this chapter.
2. Push until the SPRT type is selected.
3. Push E to show the prompt "SET(1)/RECALL(2)".
4. Push to recall the custom SPRT curve coefficients.
5. To use a different custom SPRT, push two times to show the SPRT selection
prompt.
Thermocouple (TC) Measure
The Product can measure all common thermocouple types in °F or °C, plus basic
millivolts from -10.0 to 75.0 mV.
The Product uses these common thermocouple types:
B
C
E
J
K
L
N
R
S
T
U
XK
BP
1. Disconnect test leads from external devices.
2. Push to select thermocouple and RTD/Ω mode, if not already selected. If RTD/Ω
mode is shown, push the key again to go to thermocouple mode.
3. If output mode is shown, push and to select input.