Primary Inputs and Outputs
Thermocouple (TC) Source 2
2-11
Thermocouple (TC) Source
The Product can source all common thermocouple types in °F or °C, plus basic millivolts
from -10.0 mV to 75.0 mV.
The following common thermocouple types are supported:
B
C
E
J
K
L
N
R
S
T
U
XK
BP
1. Disconnect test leads from external devices.
2. Push to select thermocouple and RTD/Ω mode, if not already selected. If RTD/Ω
mode is shown, push the key again to go to thermocouple mode.
3. If input mode is shown, push and to select output mode.
4. Push to select the necessary thermocouple type or the millivolt range.
5. Connect the UUT to the thermocouple terminals of the Product with a standard TC
miniplug as shown in Figure 2-6. One pin is wider than the other. Do not push the
plug into the incorrect polarization. The TC wire used for the connection must agree
with the thermocouple type selected for correct cold junction compensation. It is
necessary for the miniplug and terminals to become stabile to the same temperature.
They are stable in approximately 3 minutes.