Product Manual

Isolated Inputs
Voltage Input 3
3-3
Voltage Input
The Product can measure dc voltages from 0 V to 100 V. Use these ranges for maximum
accuracy: 10 V and 100 V.
1. Disconnect test leads from external devices.
2. Push to select isolated dc voltage and current input mode, if not already selected.
If the necessary dc voltage mode is not shown, push again to cycle to the
necessary dc voltage mode.
3. Connect the unit under test (UUT) to the isolated voltage/current input terminals of
the Product as in Figure 3-1.
100 V PK
MAX
HI
INPUT
LO
Switch Test
Reset
LOOP
HART
PWR
Unit Under Test (UUT)
DC Voltage Output
+
-
Press and Hold
or Switch Test
VOLTS
mA
gwp017.eps
Figure 3-1. Isolated DC Voltage Input Connection
Current Input
The Product can measure dc current from 0 mA to 50 mA.
1. Disconnect test leads from external devices.
2. Push to select isolated dc voltage and current input mode, if not already selected.
If the dc current mode is not shown, push the key again to go to it.
3. Connect the UUT to the isolated voltage/current input terminals of the Product as in
Figure 3-2.
100 V PK
MAX
HI
INPUT
LO
Switch Test
Reset
LOOP
HART
PWR
Unit Under Test (UUT)
DC Current Output
+
-
Press and Hold
or Switch Test
VOLTS
mA
gwp018.eps
Figure 3-2. Isolated DC Current Input Connection
4. If the UUT is a 2-wire loop powered transmitter that is disconnected from the wiring,