Product Manual
Application Notes
Thermocouple Transmitter 4
4-9
Thermocouple Transmitter
1. Disconnect test leads from external devices.
2. Select thermocouple output on the primary display as given in the “Thermocouple (TC)
Source” section of Chapter 4.
3. Select the correct thermocouple type for the transmitter.
4. Select current input on the isolated display as given in the “Current Input” section of
Chapter 5.
5. Select the isolated loop power option. If a HART communicator is necessary for set up of
the transmitter, select HART.
6. Connect the transmitter as shown in Figure 4-7.
7. To verify and calibrate the transmitter, refer to the transmitter documentation.