Product Manual

Application Notes
Thermocouple Transmitter 4
4-13
Precision Temperature Measurement with Probe
1. With the probe and the custom coefficients that go with the probe, the total system error
is 0.03 °C.
2. Disconnect test leads from external devices.
3. Select RTD input on the primary given as given in the “Resistance Temperature Detector
(RTD) and Ohms Measure” section of Chapter 4.
4. Select the user-defined curve with the custom coefficients for the probe. If custom
coefficients have not been recorded for the probe, follow the instructions in the
“Resistance Temperature Detector (RTD) with Custom Coefficients” section in
Chapter 4.
5. Connect the probe as shown in Figure 4-11.