Product Manual

7526A
Users Manual
6-6
Error Code List
Table 6-7 is list of error codes for the Product.
Table 6-7. Error Code List
Error
Number
Message Class
Description
1 DDE Error queue overflow.
100 EXE
OPER or STBY was received when the Calibrator is in
measure mode or thermocouple source mode.
101 CME
A non-numeric entry was received in a field that must
contain a numeric entry.
102 EXE The numeric field is more than 10 characters.
103 CME incorrect units name or prefix.
104 EXE
An attempt to go into RTD source mode was made when
SPRT is selected, or to select SPRT when RTD source
mode is selected.
105 EXE Entry is above top limit for the selected output range.
106 EXE Entry is below lower limit for the selected output range.
108 CME A necessary command parameter was missing.
109 CME
An incorrect TC_MEAS or RTD_MEAS unit parameter
(not CEL or FAR) was received, or an incorrect
PRES_UNIT or ISO_PRES_UNIT parameter was
received.
110 CME An incorrect RANGELCK parameter was received.
111 EXE
RANGELCK ON was received when the Calibrator is not
in Volts mode.
112 CME An incorrect RTD_TYPE parameter was received.
113 CME An incorrect TC_REF parameter was received.
114 CME An incorrect TSENS_TYPE parameter was received.
116 EXE
No pressure module was present when a pressure
function was requested, or the total zero offset from
calibration is more than 6%. This error can also occur if
the total zero offset from calibration is out of limits when
zeroing the thermocouple millivolts type (maximum offset
±1 mV), or the RTD ohms type (maximum ±0.1 ohm on
high range, or ±0.01 ohm on the low range).
117 CME An unrecognizable command was received.