Product Manual

Remote Commands
Remote Command List 6
6-9
FAULT?
This command shows the most recent error code from the error queue. If the queue is
empty (no errors have occurred) it shows 0. The command is typically used to verify that
the prior command did what it was intended to do.
Parameter: <None>
Response: <value>
where <value> is one of the error codes documented in
Chapter 6.
Example: FAULT?
105
This example shows the error code which will occur if the
prior command tried to set a value more than 100 mA for
current output. The error code 105 is an indications that the
value was above the maximum limit for the selected output
range.
FUNC?
This command shows the current output, measurement, or calibration function for the
primary and isolated displays.
Parameter: <None>
Response: <isolated>,<primary>
where <isolated> is one of the following:
DC10V measure DC voltage, 10V range
DC100V measure DC voltage, 100V range
DCI measure DC current
PRESSURE measure pressure
and where <primary> is one of the following:
DCV source DC voltage
DCI source DC current
RTD_OUT source RTD/Ohms
RTD_IN measure RTD/Ohms
TC_OUT source thermocouple
TC_IN measure thermocouple
PRESSURE measure pressure
Example: FUNC?
DC10V,PRESSURE
This example is an indication that the isolated display is
selected to the 10V range and the primary display is selected
to pressure.