2 Fluke Corporation The Fluke 568 Ex Intrinsically Safe Infrared Thermometer
Fluke. The Most Trusted Tools
in the World.
Specifications
Ordering information
FLUKE-568 Ex Intrinsically Safe Infrared Thermometer
Included equipment
K-type thermocouple bead probe
Conductive IS hard carrying case
Users manual
Safety Certifications
Agency Safety rating
ATEX/IECEx
Zone 1 and 2
IECEx EPS 13.0006X
Ex ia IIC T4 Gb
0 °C ≤ Ta ≤ 50 °C
EPS 13 ATEX 1.525 X
II 2G Ex ia IIC T4 Gb
NEC-500/NEC-505 (pending)
Class I Division 1 and 2
Class I, Division 1, Groups
ABCD T4
Class I, Division 2, Groups
ABCD T4
Class I, Zone 1, AEx ia IIC T4
Ex ia IIC T4
0 °C ≤ Ta ≤ 50 °C
GOST (pending)
Zone 1 and 2
POCC DE.
ГБ05.B
Ex ia IIC T4 Gb X
ОТ 0 °C ДО +50 °C
EPS 13 ATEX 1 525 X
II 2G Ex ia IIC T4 Gb
0 °C ≤ Ta ≤ 50 °C
PCEC (pending)
Zone 1 and 2
PCEC
Ex ia IIC T4 Gb
CE13.
EPS 13 ATEX 1 525 X
II 2G Ex ia IIC T4 Gb
0 °C ≤ Ta ≤ 50 °C
INMETRO (pending)
Zone 1 and 2
IEx 13.0122X
Ex ia IIC T4 Gb
EPS 13 ATEX 1 525 X
II 2G Ex ia IIC T4 Gb
0 °C ≤ Ta ≤ 50 °C
Fluke 568 Ex Infrared Thermometer
Infrared temperature
range
-40 °C to 800 °C (-40 °F to 1472 °F)
Infrared accuracy < 0 °C (32 °F): ± (1.0 °C (± 2.0 °F) + 0.1°/1 °C or °F);
≥ 0 °C (32 °F): ± 1 % or ± 1.0 °C (± 2.0 °F),
whichever is greater
Display resolution 0.1 °C/0.1 °F
Infrared spectral response 8 µm to 14 µm
Infrared response time < 500 ms
Thermocouple Type-K
input temperature range
-270 °C to 1372 °C (-454 °F to 2501 °F)
Thermocouple Type-K
input accuracy
<-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is greater
<-40 °F: ±(2 °F + 0.2 °/1 °F)
≥-40 °F: ±1 % or 2 °F, whichever is greater
D:S (distance to measure-
ment spot size)
50:1
Laser sighting Single-point laser
Minimum spot size 19 mm (0.75 in)
Emissivity adjustment By built-in table of common materials or digitally
adjustable from 0.10 to 1.00 by 0.01
Data storage 99 points
Hi/Low alarms Audible and two-color visual
Min/Max/Avg/Dif Yes
Display Dot matrix with function menus
Backlight Two levels, normal and extra bright for darker
environments
Trigger lock Yes
Switchable Celsius and
Fahrenheit
Yes
Power 2 AAA/LR03 type-approved Batteries (For a list of
type-approved batteries, refer to Product Safety
Instructions)
Battery life 4 hours with laser and backlight on; 100 hours with
laser and backlight off, at 100 % duty cycle
Operating temperature 0 °C to 50 °C (32 °F to 122 °F)
Storage temperature -20 °C to 60 °C (-40 °F to 140 °F)
Bead thermocouple
Type-K range
-40 °C to 260 °C (-40 °F to 500 °F)
Bead thermocouple
Type-K accuracy
± 1.1 °C (2.0 °F) from 0 °C to 260 °C
(32 °F to 500 °F), typically within 1.1 °C (2.0 °F)
from -40 °C to 0 °C (-40 °F to 32 °F)