Data Sheet

3 Fluke Corporation Fluke 430 Series II Three-Phase Energy and Power Quality Analyzers
Technical specifications
Specifications are valid for models Fluke 434-II, Fluke 435-II, Fluke 437-II unless otherwise specified.
Specificatons for Amp and Watt readings are based upon i430-Flexi-TF unless otherwise specified.
Input characteristics
Voltage inputs
Number of inputs 4 (3 phase + neutral) dc-coupled
Maximum input voltage 1000 Vrms
Nominal voltage range Selectable 1 V to 1000 V
Max. peak measurement voltage 6 kV (transient mode only)
Input impedance 4 MΩ//5 pF
Bandwidth > 10 kHz, up to 100 kHz for transient mode
Scaling 1:1, 10:1, 100:1, 1,000:1 10,000:1 and variable
Current inputs
Number of inputs 4 (3 phase + neutral) dc- or ac-coupled
Type Clamp or current transformer with mV output or i430flex-TF
Range 0.5 Arms to 600 Arms with included i430flex-TF (with sensitivity 10x)
5 Arms to 6000 Arms with included i430flex-TF (with sensitivity 1x)
0.1 mV/A to 1 V/A and custom for use with optional ac or dc clamps
Input impedance 1 MΩ
Bandwidth > 10 kHz
Scaling 1:1, 10:1, 100:1, 1,000:1 10,000:1 and variable
430 Series II Power Quality and Energy Analyzer selection table
Model Fluke 434-II Fluke 435-II Fluke 437-II
Standard compliance IEC 61000-4-30 Class S IEC 61000-4-30 Class A IEC 61000-4-30 Class A
Volt Amp Hz
Dips and swells
Harmonics
Power and energy
Energy loss calculator
Unbalance
Monitor
Inrush
Event waveform capture
Flicker
Transients
Mains signaling
Power wave
Power inverter efficiency
400Hz
C1740 Soft Case
C437-II Hard Case with rollers
SD card (Max 32 GB) 8 GB 8 GB 8 GB
All models include the following accessories TL430 test lead set, 4 x i430 thin flexi current probes, BP290 battery, BC430 power adapter with international power adapter set, USB cable
A-B mini and PowerLog CD.