Performance Verification
4.8 Sags & Swells 4
4-29
3. Press to move the right cursor to the fifth division.
4. Observe the readings and verify that they are between 1.38 and 1.62 kA.
5. When you are finished, set the 5500A to Standby
4.8 Sags & Swells
Proceed as follows to test the sags & swells mode.
ST8001.CGM
Figure 4-9. Test Tool Input 1 & 2 5500A Normal Output
1. Connect the test tool to the 5500A as shown in Figure 4.9.
2. Press
.
3. Press
to highlight SAGS & SWELLS.
4. Press
to enter mode.
Now the RECORD TIME is highlighted. If the time is not 4 minutes then:
1. Press
2. Press
to highlight 4 minutes.
3. Press
to confirm; mark changes to .
4. Set the 5500A to source a sine wave of 5V, 60Hz (NORMAL output, MODE wave
sine).
5. Press
to highlight START.
6. Press
7. After approximately 30 seconds press
.
8. Press
and move the cursor into the measured region.
9. Check the readings MAX, V
,MIN. of Input 1 is between 4.80 and 5.20.
10. Check the readings MAX, A
,MIN. of Input 2 is 4.80 and 5.20.
11. When you are finished, set the 5500A to Standby