v3001 FC Wireless DC Voltage Calibration Manual July 2014 © 2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days.
Table of Contents Title Page Introduction ........................................................................................................ Contact Fluke ..................................................................................................... Safety Information .............................................................................................. Symbols .............................................................................................................
v3001 FC Calibration Manual ii
List of Tables Table 1. 2. 3. 4. 5. Title Symbols .................................................................................................................. Required Equipment ............................................................................................... Performance Tests ................................................................................................. Calibration Adjustment ...........................................................................................
v3001 FC Calibration Manual iv
List of Figures Figure Title 1. 2. 3. 4. All Segments of the Display ................................................................................... Performance Test Connections .............................................................................. Calibration Password Reset ................................................................................... Battery Replacement ..............................................................................................
v3001 FC Calibration Manual vi
Introduction Warning Read "Safety Information" before you use the Product. This manual has the verification and calibration adjustment procedures for the v3001 FC Wireless DC Voltage module (the Product). Please see the v3001 FC Quick Reference Guide for usage information.
v3001 FC Calibration Manual Safety Information A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test. Warning To prevent possible electrical shock, fire, or personal injury: 2 • Carefully read all instructions. • Use the Product only as specified, or the protection supplied by the Product can be compromised.
Wireless DC Voltage Symbols • Connect the common test lead before the live test lead and remove the live test lead before the common test lead. • Remove the input signals before you clean the Product. • Do not operate the Product with covers removed or the case open. Hazardous voltage exposure is possible. For safe operation and maintenance of the Product: • Remove batteries to prevent battery leakage and damage to the Product if it is not used for an extended period.
v3001 FC Calibration Manual Specifications Accuracy is specified as ±( [% of reading] + [number of least significant digits] ). All ranges are autoranging. Accuracy is specified from 1 % to 100 % of the range obtained by autoranging, from 18 °C to 28 °C. DC Voltage Range 600.0 mV 6.000 V 60.00 V 600.0 V 1000 V Resolution 0.1 mV 0.001 V 0.01 V 0.1 V 1V Accuracy ± [0.09 % + 3 digits] ± [0.15 % + 2 digits] General LCD w/Backlight....................................................
Wireless DC Voltage Performance Tests Required Equipment The equipment in Table 2 is necessary for performance tests and calibration adjustment. Table 2. Required Equipment Recommended Equipment Fluke 5522A Calibrator (or equivalent) Function DC Volts Required Characteristics 6 mV to 1000 V, ±0.02 % Performance Tests Warning To prevent possible electrical shock, fire, or personal injury, do not perform the performance test procedures unless the Product is fully assembled.
v3001 FC Calibration Manual Backlight To verify that the backlight functions: 1. With the Product on, push . 2. The backlight will come on. If it does not, repair is necessary. See “Contact Fluke”. Keypad Test To verify that the keypad functions, turn on the Product and push each button separately. Each button push will turn on a display annunciator. If the buttons do nothing, repair is necessary. See “Contact Fluke”. DC Voltage Test Before you do the dc voltage test: 1.
Wireless DC Voltage Before Calibration Adjustment Figure 2. Performance Test Connections hcc07.eps Before Calibration Adjustment Before the Product calibration can be adjusted, you must go through the maintenance mode menu and enter your password. Maintenance Mode The Product maintenance mode can be used to set different parameters on the Product that include auto power off, backlight adjustment, and calibration. To use the maintenance mode: 1. With the Product on, push and hold . 2. Push .
v3001 FC Calibration Manual Password Entry To go to the calibration mode, push five times until CAL is shown. You will need to enter a password to access calibration mode. To enter the password: 1. Push and the CAL counter is shown. For example n002. 2. Push to show “????”. 3. Push to change the first digit of your password (default: 1234). 4. Push to confirm your choice. The subsequent “?” changes to “_”. 5. Do steps 3 and 4 again to enter the subsequent digits of the four-number password. 6.
Wireless DC Voltage Before Calibration Adjustment Restore the Default Password If the calibration password is lost, the default password (1234) can be manually restored with the subsequent steps: Warning To prevent electric shock or personal injury, remove all input signals before you open the Product. 1. Remove the Product battery door. See “Battery Replacement”. 2. With a Phillips screwdriver, remove the bottom case screws. Two of the screws are inside of the battery door. 3.
v3001 FC Calibration Manual Calibration Adjustment The Product features closed-case calibration adjustment and uses known reference sources. The Product measures the applied reference source, calculates correction factors, and stores the correction factors in nonvolatile memory. Should the Product fail any of the performance tests, do the calibration adjustment procedure. When “C-01” is shown on the display, apply the correct input signal shown in Table 4 to the Product.
Wireless DC Voltage Maintenance Maintenance Clean the Product Caution To prevent possible damage to the Product or to equipment under test, do not use abrasive cleaners. They will damage the case. To clean the Product, use a cloth with a mild cleaning solution. Battery Replacement Warning To prevent possible explosion, fire, or personal injury, replace the batteries when the low battery indicator () shows to prevent incorrect measurements.
v3001 FC Calibration Manual Figure 4. Battery Replacement hcc001.eps User-Replaceable Parts User-replaceable parts are shown in Table 5. Table 5. User-Replaceable Parts Fluke Part Number Description 4130305 FLK-3000-2003, DOOR, BATTERY 1 1881997 TPAK Meter Hanging Kit 1 variable[1] TEST LEADS 2 variable LARGE ALLIGATOR CLIP BLACK, BULK 1 variable[1] LARGE ALLIGATOR CLIP RED, BULK 1 376756 Battery, AA 1.