Product Manual

Before You Start
Optional Accessories
1
1-7
Optional Accessories
ac adapter with 14-pin LEMO connector
replacement lenses, macro attachments
protective window for lenses, laser protection filter
Bluetooth headset
gigabit Ethernet
Ethernet cable with 8-pin LEMO connector
device tripod
spare batteries
Technical Description
Functional Principle
The Product is a thermography system designed for the
long-wave infrared spectral range (LWIR) of 7.5 μm to
14 μm. The lens reproduces the object scene to a
microbolometer array with 1024 x 768 and/or 640 x 480
pixels. The electrical signal of the detector array is
processed further by the internal electronics. In this, the
electronics comprise all functions required for camera
operation, such as actuation of the microbolometer array,
A/D conversion, offset and gain correction, pixel
correction, as well as actuation of the different interfaces.
Operation is optionally performed directly using the
integrated keyboard (joystick, focus rocker switch,
function buttons) or using optional interfaces. RS232,
Ethernet (GigE Vision), or USB2.0 are available as
interfaces. Specific software packages are required in
order to use these interfaces.
Description of the Functional Units
Lens
The camera lens collects the IR radiation in the field of
view emitted by the measuring object and reproduces this
IR radiation on the detector array. Field of view (FOV)
and resolution/measuring spot size (IFOV, Instantaneous
Field of View) are determined at the same distance to the
focal distance f of the used lens.